Frequency-domain optical tomographic imaging of arthritic finger joints.
Where this comes from
- Record sourced from PubMed, PMID 21964730.
- Also identified by DOI 10.1109/TMI.2011.2135374.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We are presenting data from the largest clinical trial on optical tomographic imaging of finger joints to date. Overall we evaluated 99 fingers of patients affected by rheumatoid arthritis (RA) and 120 fingers from healthy volunteers. Using frequency-domain imaging techniques we show that sensitivities and specificities of 0.85 and higher can be achieved in detecting RA. This is accomplished by deriving multiple optical parameters from the optical tomographic images and combining them for the statistical analysis. Parameters derived from the scattering coefficient perform slightly better than absorption derived parameters. Furthermore we found that data obtained at 600 MHz leads to better classification results than data obtained at 0 or 300 MHz.
Medical subject headings
- Arthritis, Rheumatoid
- Diagnosis, Computer-Assisted
- Finger Joint
- Tomography, Optical