Scaling digital radiographs for templating in total hip arthroplasty using conventional acetate templates independent of calibration markers.

Brew, Christopher J; Simpson, Philip M; Whitehouse, Sarah L; Donnelly, William; Crawford, Ross W; Hubble, Matthew J W · J Arthroplasty · 2012

case_series · Level IV

Where this comes from

Abstract

We describe a scaling method for templating digital radiographs using conventional acetate templates independent of template magnification without the need for a calibration marker. The mean magnification factor for the radiology department was determined (119.8%; range, 117%-123.4%). This fixed magnification factor was used to scale the radiographs by the method described. Thirty-two femoral heads on postoperative total hip arthroplasty radiographs were then measured and compared with the actual size. The mean absolute accuracy was within 0.5% of actual head size (range, 0%-3%) with a mean absolute difference of 0.16 mm (range, 0-1 mm; SD, 0.26 mm). Intraclass correlation coefficient showed excellent reliability for both interobserver and intraobserver measurements with intraclass correlation coefficient scores of 0.993 (95% CI, 0.988-0.996) for interobserver measurements and intraobserver measurements ranging between 0.990 and 0.993 (95% CI, 0.980-0.997).