Use of a wire extender during neuroprotected vertebral artery angioplasty and stenting.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 21990622.
- Also identified by DOI 10.1136/jnis.2010.002311.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The off-label use of an extender wire during vertebral artery stenting and angioplasty with or with neuroprotection has not been previously reported. Retrospective, single-patient, technical report. After monorail balloon angioplasty was performed on a proximal left vertebral artery stenosis, the 190 cm long Accunet neuroprotection filter device was not long enough for delivery of an over-the-wire stent. After mating a 145 cm long, 0.014 inch extension wire to the filter device, a balloon-mounted Liberté stent was implanted with good angiographic and clinical results. The off-label use of an extender wire permits successful over-the-wire stenting on a monorail neuroprotection device for vertebral artery endosurgery.
Medical subject headings
- Angioplasty, Balloon
- Embolic Protection Devices
- Stents
- Vertebral Artery