Laboratory evaluation and use of the Engström metabolic computer in the clinical setting.
Where this comes from
- Record sourced from PubMed, PMID 2199150.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Laboratory evaluation and use of the Engström metabolic computer (EMC) in a clinical setting are described. The accuracy of the EMC was tested using an inert gas dilution technique. Mean errors in oxygen consumption (VO2) and CO2 production compared with predicted values were less than 2% of predicted, and generally less than 1%, but with an SD of less than or equal to 5.1%. At an FIO2 of greater than or equal to 0.7, the errors in VO2 were erratic and generally greater than 15%. The apparatus has the facility to assume an RQ in the calculation of VO2 and the errors in measured VO2 were less than 1% up to an FIO2 of 0.8.
Medical subject headings
- Computers
- Energy Metabolism