Embol-X intraaortic filter and transaortic approach for improved cerebral protection in transcatheter aortic valve implantation.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 22051319.
- Also identified by DOI 10.1016/j.athoracsur.2011.05.109.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Techniques for transcatheter aortic valve implantation involve manipulation of the native aortic valve and the aortic arch. As a result, excellent technical outcomes of this technique remain counterbalanced by a level of neurologic complications that remain unacceptably high. We present here a report of a new potential combined approach to reduce periprocedural neurologic events. After a 6-cm J-shaped upper-mini-sternotomy, 26-mm Sapien valve (Edwards Lifesciences, Irvine, CA) was deployed through a transaortic approach. In addition, the Embol-X device (Edwards Lifescience) was directly inserted in the distal ascending aorta was deployed during all the procedure. Postoperative evolution was clinically uneventful. Postoperative magnetic resonance imaging revealed no new ischemic lesions in this first patient.
Medical subject headings
- Aortic Valve
- Embolic Protection Devices
- Heart Valve Prosthesis
- Heart Valve Prosthesis Implantation
- Intracranial Embolism
- Intracranial Thrombosis