Respiratory function monitor guidance of mask ventilation in the delivery room: a feasibility study.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 22056350.
- Also identified by DOI 10.1016/j.jpeds.2011.09.017.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To investigate whether using a respiratory function monitor (RFM) during mask resuscitation of preterm infants reduces face mask leak and improves tidal volume (V(T)). Infants receiving mask resuscitation were randomized to have the display of an RFM (airway pressure, flow, and V(T) waves) either visible or masked. Twenty-six infants had the RFM visible, and 23 had the RFM masked. The median mask leak was 37% (IQR, 21%-54%) in the visible RFM group and 54% (IQR, 37%-82%) in the masked RFM group (P = .01). Mask repositioning was done in 19 infants (73%) of the visible group and in 6 infants (26%) of the masked group (P = .001). The median expired V(T) was similar in the 2 groups. Oxygen was provided to 61% of the visible RFM group and 87% of the RFM masked group (P = .044). Continuous positive airway pressure use was greater in the visible RFM group (73% vs 43%; P = .035). Intubation in the delivery room was done in 21% of the visible group and in 57% of the masked group (P = .035). Using an RFM was associated with significantly less mask leak, more mask adjustments, and a lower rate of excessive V(T).
Medical subject headings
- Delivery Rooms
- Infant, Premature
- Masks
- Monitoring, Physiologic
- Positive-Pressure Respiration
- Respiration
- Resuscitation