Ulnar component surface finish influenced the outcome of primary Coonrad-Morrey total elbow arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 22104128.
- Also identified by DOI 10.1016/j.jse.2011.08.062.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Controversy remains regarding the mechanisms leading to ulnar loosening after elbow replacement. We therefore assessed the loosening rate of the ulnar component as a function of the surface finish of a commonly used implant design. This study included 3 groups who received implants with 3 different surface finishes: sintered beads (278 components), polymethylmethacrylate precoating (219 components), and plasma-sprayed beads (205 components). The 3 groups who received the implants did not differ statistically in age, sex, or underlying diagnosis. The rates of mechanical failure for the sintered, precoated, and plasma-sprayed ulnar components were 6.8%, 12.8%, and 0%, respectively. The 7-year Kaplan-Meier survival rates free of mechanical failure were 93.1% for the sintered-beads group, 83.1% for the precoated group, and 100% for the plasma-sprayed group. Failed precoated ulnar components often failed early and exhibited typical features, including proximal debonding and severe focal or global osteolysis, sometimes leading to periprosthetic insufficiency fractures. Precoating the ulnar component with polymethylmethacrylate can lead to an increased rate of loosening and severe osteolysis. A plasma-sprayed ulnar component is associated with a very low mechanical failure rate with surveillance of less than 10 years.
Medical subject headings
- Arthroplasty, Replacement, Elbow
- Elbow Prosthesis
- Prosthesis Failure
Anatomy
- elbow
- ulna