An arteriovenous fistula model of intimal hyperplasia for evaluation of a nitinol U-Clip anastomosis.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 22104322.
- Also identified by DOI 10.1016/j.ejvs.2011.11.002.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to create an ovine arteriovenous fistula (AVF) model which would closely replicate a human forearm fistula and use this to quantify the degree of intimal hyperplasia in those created with the U-Clip compared to a conventional sutured anastomosis. Twenty AVFs were created in 10 Border Leicester-Merino sheep between the superficial femoral artery and vein of each hind limb. On one side the U-Clip and on the other a continuous polypropylene suture was used to perform the anastomosis. The animals were sacrificed at 2 (n = 3), 4 (n = 4), 6 (n = 3) weeks and histological slices were taken of each AVF in cross section to determine the intimal media area per unit length (IMA/L). Intimal hyperplasia (IH) was observed at all time points with one AVF found occluded with thrombus at the time of harvest. The IMA/L was significantly lower in the U-Clip groups by 24% at 2 weeks, 32% at 4 weeks and 23% at 6 weeks (Two-way ANOVA, p = 0.019, observed power = 0.825, time or side p ≥ 0.766, type p = 0.001; Paired t-test, p < 0.001 between matched anastomotic types). Time taken to perform the anastomosis was similar between the two anastomotic techniques (Polypropylene 14(8-18) vs. U-Clip 15.3(11-23) min; p = 0.47). This ovine AVF model results in IH similar to that seen in a human AVF. The IH that occurs with the U-Clip is less than that of continuous polypropylene suture.
Medical subject headings
- Alloys
- Anastomosis, Surgical
- Arteriovenous Fistula
- Surgical Instruments
- Sutures
- Tunica Intima