A nondivergent three field matching technique for breast irradiation.

Chu, J C; Solin, L J; Hwang, C C; Fowble, B; Hanks, G E; Goodman, R L · Int J Radiat Oncol Biol Phys · 1990

other · Level V

Where this comes from

Abstract

Effective radiation therapy to intact breasts requires the delivery of adequate dose to a large target volume using complex beam arrangements. A semi-empirical method is described to determine the correct gantry, collimator, and couch positions for a geometrically accurate field match among adjoining radiation beams. The technique uses a metal rod and chain combination to aid determination of the proper couch setting under remote fluoroscopy control. A couch position error of more than half a degree is easily detectable by this technique.

Medical subject headings