Automated algorithm for generalized tonic-clonic epileptic seizure onset detection based on sEMG zero-crossing rate.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 22156944.
- Also identified by DOI 10.1109/TBME.2011.2178094.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Patients are not able to call for help during a generalized tonic-clonic epileptic seizure. Our objective was to develop a robust generic algorithm for automatic detection of tonic-clonic seizures, based on surface electromyography (sEMG) signals suitable for a portable device. Twenty-two seizures were analyzed from 11 consecutive patients. Our method is based on a high-pass filtering with a cutoff at 150 Hz, and monitoring a count of zero crossings with a hysteresis of ±50 μV . Based on data from one sEMG electrode (on the deltoid muscle), we achieved a sensitivity of 100% with a mean detection latency of 13.7 s, while the rate of false detection was limited to 1 false alarm per 24 h. The overall performance of the presented generic algorithm is adequate for clinical implementation.
Medical subject headings
- Algorithms
- Diagnosis, Computer-Assisted
- Electromyography
- Epilepsy, Tonic-Clonic
- Seizures
- Signal Processing, Computer-Assisted