Catheter tracking with phase information in a magnetic resonance scanner.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 22186949.
- Also identified by DOI 10.1109/TMI.2011.2179944.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study is to describe a new active technique for accurately determining both the position and orientation of the tip of a catheter during magnetic resonance (MR)-guided percutaneous cardiovascular procedures. The technique utilizes phase information introduced into the MR signal from a small receive coil located on the distal tip of the catheter. Phase patterns around a small receive coil are rich in information that is directly related to position and orientation. This information can be collected over a large spherical volume with a diameter several times that of the receive coil. The high degree of redundancy yields the potential for an accurate and robust method of catheter tracking. A tracking algorithm is presented that performs catheter tip localization using phase images acquired in two orthogonal planes without any a priori knowledge of catheter position. Associated experimentation demonstrating feasibility is also presented.
Medical subject headings
- Cardiac Catheterization
- Magnetic Resonance Imaging, Interventional
- Magnetics
- Pattern Recognition, Automated