Accuracy and responsiveness of the stepwatch activity monitor and ActivPAL in patients with COPD when walking with and without a rollator.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 22200080.
- Also identified by DOI 10.3109/09638288.2011.641666.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the measurement properties of the StepWatch(™) Activity Monitor (SAM) and ActivPAL in COPD. Whilst wearing both monitors, participants performed walking tasks at two self-selected speeds, with and without a rollator. Steps obtained using the monitors were compared with that measured by direct observation. Twenty participants aged 73 ± 9 years (FEV(1) = 35 ± 13% pred; 8 males) completed the study. Average speeds for the slow and normal walking tasks were 34 ± 7 m·min(-1)and 46 ± 10 m·min(-1), respectively. Agreement between steps recorded by the SAM with steps counted was similar irrespective of speed or rollator use (p = 0.63) with a mean difference and limit of agreement (LOA) of 2 steps·min(-1) and 6 steps·min(-1), respectively. Agreement for the ActivPAL was worse at slow speeds (mean difference 7 steps·min(-1); LOA 10 steps·min(-1)) compared with normal speeds (mean difference 4 steps·min(-1); LOA 5 steps·min(-1)) (p = 0.03), but was unaffected by rollator use. The change in step rate between slow and normal walking via direct observation was 12 ± 7 steps·min(-1) which was similar to that detected by the SAM (12 ± 6 steps·min(-1)) and ActivPAL (14 ± 7 steps·min(-1)). The SAM can be used to detect steps in people who walk very slowly including those who use a rollator. Both devices were sensitive to small changes.
Medical subject headings
- Ergometry
- Monitoring, Ambulatory
- Pulmonary Disease, Chronic Obstructive
- Walkers
- Walking