Measurement of entrainment ratio during high frequency jet ventilation.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 2223336.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We have measured tidal (VT), entrained (Ve) and "blowback" (Vbb) volumes during high frequency jet ventilation (HFJV) through a Mallinckrodt Hi-Lo Jet tracheal tube in anaesthetized patients. The above volumes were calculated by digital integration of the appropriate regions of flow curves derived from a pneumotachograph placed between the bias flow tubing and the tracheal tube. At a driving pressure of 1 bar, lung minute ventilation increased with increasing ventilatory frequency, whilst tidal volumes decreased. The contribution of entrainment to tidal volume (Ve/VT) remained constant, although the volumes entrained were relatively small and varied widely from subject to subject. Blowback volumes were considerable, especially at ventilatory frequencies used clinically (1-2 Hz). We conclude that it is not possible to entrain predictable concentrations of volatile agents from the low pressure bias flow during HFJV.
Medical subject headings
- High-Frequency Jet Ventilation
- Respiration