Endovascular retrieval of a dislodged femoral arterial closure device with Alligator forceps.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 22265797.
- Also identified by DOI 10.1016/j.jvs.2011.12.015.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Angio-Seal (St. Jude Medical, Minnetonka, Minn) is a frequently used percutaneous femoral artery closure device to obtain hemostasis after arterial catheterization. Separation and migration of the device, leading to distal embolization and limb-threatening ischemia necessitating emergency surgery, is reported but extremely rare. We present a unique case of successful endovascular bailout management of a dislocated Angio-Seal with use of an Alligator Tooth Retrieval forceps (Cook Medical, London, United Kingdom).
Medical subject headings
- Angioplasty
- Equipment Failure
- Femoral Artery
- Hemostatic Techniques
- Ischemia
- Leg