Effect of femoral component design on patellofemoral crepitance and patella clunk syndrome after posterior-stabilized total knee arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 22285232.
- Also identified by DOI 10.1016/j.arth.2011.12.009.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine if recent changes to the femoral component of a particular posterior-stabilized total knee prosthesis would affect the incidence of postoperative patellofemoral crepitance and patella clunk syndrome. One hundred eight total knee arthroplasties were performed with the conventional design; 136 were performed after the femoral component was changed. Complications were compared between the groups with an average follow-up of 17.7 months and 12.4 months, respectively. Thirteen knees with the conventional design (12%) were found to have patellofemoral complications; no complications were noted with the new design (P < .0001). Femoral components with a deep trochlear groove and smooth transition of the intercondylar box appear to better accommodate any peripatellar fibrous nodule that may form after total knee arthroplasty.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Femur
- Knee Prosthesis
- Patellofemoral Pain Syndrome
- Postoperative Complications
- Prosthesis Design
- Prosthesis Failure
Anatomy
- femur
- knee