Measurement of anterior-posterior displacement of the knee. A comparison of the results with instrumented devices and with clinical examination.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 2229105.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Thirteen subjects who had normal knees and fifteen patients who had a chronic rupture of the anterior cruciate ligament were tested in order to compare the measurements of tibiofemoral displacement as recorded by four commercial devices: the Acufex knee-signature system, the Genucom knee-analysis system, the Medmetric KT-1000 arthrometer, and the Stryker knee-laxity tester. Anterior and posterior displacement were measured at forces of eighty-nine newtons (twenty pounds) and 133 newtons (thirty pounds). We found significant differences in reproducibility of measurement among the devices. The Acufex, Medmetric, and Stryker devices had more reproducible measurements, and they could be used to identify 80 to 90 per cent of the normal subjects and anterior cruciate-deficient patients. The Genucom device had poorer reproducibility of measurement, and it tended to register greater differences in displacement between the right and left knees of normal subjects.
Medical subject headings
- Anterior Cruciate Ligament Injuries
- Joint Instability
- Knee Joint
- Orthopedic Equipment
Anatomy
- knee