Threaded acetabular components for primary and revision total hip arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 2230817.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A clinical and radiographic review of 48 total hip arthroplasty patients with threaded acetabular components was undertaken at 24-44 months of follow-up study. Twenty-five patients had primary hip arthroplasties and 23 had revision procedures. Clinical scores revealed good to excellent results in 60% of primary and 30% of revision procedures. Radiographic analysis revealed stable acetabular components in 88% of primary and 61% of revision procedures. Potentially loose acetabular components were noted in 8% of primary and 4.3% of revision procedures and loose acetabular components in 4% of primary and 34.7% of revision procedures. The rate of acetabular component loosening was considered unacceptably high in revision cases and an area of concern in primary cases. Discretionary use of these components is advised.