Ultrasound waves, widely used as a non-invasive diagnostic modality, were recently shown to stimulate neuronal activity. Functionally meaningful stimulation, as is required in order to form a unified percept, requires the dynamic generation of simultaneous stimulation patterns. In this paper, we examine the general feasibility and properties of an acoustic retinal prosthesis, a new vision restoration strategy that will combine ultrasonic neuro-stimulation and ultrasonic field sculpting technology towards non-invasive artificial stimulation of surviving neurons in a degenerating retina. We explain the conceptual framework for such a device, study its feasibility in an in vivo ultrasonic retinal stimulation study and discuss the associated design considerations and tradeoffs. Finally, we simulate and experimentally validate a new holographic method--the angular spectrum-GSW--for efficient generation of uniform and accurate continuous ultrasound patterns. This method provides a powerful, flexible solution to the problem of projecting complex acoustic images onto structures like the retina.
Medical subject headings
- Prosthesis Design
- Retina
- Ultrasonics