Use of the Ascent balloon for a 2-in-1 remodeling technique: feasibility and initial experience: case report.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 22343831.
- Also identified by DOI 10.1227/NEU.0b013e31822c49ad.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To present the feasibility of using the Ascent balloon, a new double-lumen remodeling balloon, for a new 2-in-1 technique allowing coiling through the lumen of the balloon without the use of an additional coiling microcatheter. Remodeling technique had enlarged the indications for endovascular treatment of intracranial aneurysm. Nevertheless, one of the limitations of this technique is that it requires using 2 devices in the same parent artery. A 55-year-old woman presented with a 7.7 × 4.5-mm incidental anterior communicating artery aneurysm. Only 1 A1 segment (left side) was patent on the cerebral angiogram. A 6F Fargo Max guiding catheter was positioned in the left petrous internal carotid artery. The Ascent balloon was placed in front of the neck of the aneurysm after navigation on a Traxcess 0.014-in guidewire. Coiling of the aneurysm sac was performed via 1 lumen of the device under iterative inflations of the balloon through the second lumen. This new 2-in-1 technique using a sole remodeling balloon without an additional coiling microcatheter is very promising, especially in cases of a small-caliber parent artery.
Medical subject headings
- Angioplasty, Balloon
- Balloon Occlusion
- Catheterization
- Intracranial Aneurysm
- Prosthesis Implantation