Response time of nanofluidic electrochemical sensors.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 22361835.
- Also identified by DOI 10.1039/c2lc21104a.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Nanofluidic thin-layer cells count among the most sensitive electrochemical sensors built to date. Here we study both experimentally and theoretically the factors that limit the response time of these sensors. We find that the key limiting factor is reversible adsorption of the analyte molecules to the surfaces of the nanofluidic system, a direct consequence of its high surface-to-volume ratio. Our results suggest several means of improving the response time of the sensor, including optimizing the device geometry and tuning the electrode biasing scheme so as to minimize adsorption.
Medical subject headings
- Electrochemical Techniques
- Microfluidic Analytical Techniques