Treatment of aortic arch aneurysms with a modular transfemoral multibranched stent graft: initial experience.

Lioupis, C; Corriveau, M-M; MacKenzie, K S; Obrand, D I; Steinmetz, O K; Abraham, C Z · Eur J Vasc Endovasc Surg · 2012

case_series · Level IV

Where this comes from

Abstract

To present initial experience with a new modular transfemoral multibranched stent graft for treating aortic arch aneurysms. Six patients, considered high risk for open surgery, were treated with custom made branched stent grafts. All patients had a staged left carotid subclavian bypass before the endovascular procedure. Each branched graft had a 12 mm side branch for the innominate artery and an 8 mm side branch for the left common carotid artery. Four patients out of six had uneventful placement of the prostheses, with successful exclusion of their aneurysms. One patient developed a type I endoleak that was managed successfully with coiling and gluing of the aneurysm sac. In one patient, cannulation of the innominate branch was unsuccessful and an extra-anatomic bypass was necessary to perfuse the right carotid and vertebral arteries. This patient developed a stroke, while one more suffered a right cerebellar infarct. We have demonstrated the technical feasibility of a modular transfemoral branched stent graft for treatment of aortic arch aneurysms. The method is relatively safe based on initial experience. More cases and long-term follow up are necessary to evaluate the efficacy and safety of this new device.

Medical subject headings