Cutaneous manifestation of metallosis in a metal-on-metal total hip arthroplasty after acetabular liner dissociation.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 22397858.
- Also identified by DOI 10.1016/j.arth.2012.01.008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this case report, we describe a cutaneous manifestation of extensive metallosis in a patient 4 months post-metal-on-metal total hip arthroplasty with a Pinnacle cup with dissociation of the liner from the shell and resultant stripe burnishing of the shell and notch wear of the femoral neck. Dissociation of a metal liner has not been previously reported with this implant. Cutaneous metallosis has only been reported once in the literature. Clinicians should heighten their suspicion for metallosis secondary to hardware failure when encountering patients with skin discoloration in the setting of a painful and poorly functioning hip arthroplasty. In patients with failure of a metal-on-metal prosthesis with a modular metal liner in the acetabular component, liner dissociation must be considered.
Medical subject headings
- Hip Prosthesis
- Metal-on-Metal Joint Prostheses
- Poisoning
- Prosthesis Failure
- Skin Diseases
Anatomy
- hip
- pelvis