Another Selectatec switch malfunction.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 2240537.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Malfunction of a Selectatec-3 switch mechanism is reported which caused isolation of the selected vaporizer. This resulted in a delivery of unsupplemented fresh gas to the patient.
Medical subject headings
- Anesthesiology
- Nebulizers and Vaporizers