Transtelephonic defibrillation.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 2240746.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study was undertaken to assess the safety and reliability of a device for transtelephonic defibrillation. The transtelephonic system consists of a patient unit and a base station. The patient unit contains a monitor-defibrillator, electrode pads, microphone, microprocessor, and DC defibrillator. The base station comprises a control panel, computer, and ECG display. Fifteen patients who were treated in our emergency department for cardiac arrest were placed on patient units that activated our base station in a remote location within the ED. Thirteen patients were treated for ventricular fibrillation, and two patients were treated for ventricular tachycardia. Thirty-one shocks were delivered transtelephonically. In all cases, voice and ECG transmission were established without difficulty. We conclude that this system represents a safe and reliable method for the treatment of ventricular fibrillation, and we advocate additional use to study the prehospital applications of transtelephonic defibrillation.
Medical subject headings
- Electric Countershock
- Heart Arrest
- Telephone