A randomised comparison of the LMA SupremeTM and LMAProSealTM in children.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 22420717.
- Also identified by DOI 10.1111/j.1365-2044.2012.07088.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We conducted a randomised trial comparing the size-2 LMA Supreme™ with the LMA ProSeal™ in 60 children undergoing surgery. The outcomes measured were airway leak pressure, ease and time for insertion, fibreoptic examination, incidence of gastric insufflation, ease of gastric tube placement, quality of the airway during anaesthetic maintenance and complications. There were no statistically significant differences between the LMA Supreme and LMA ProSeal in median (IQR [range]) insertion time (12 (10-15 [7-18]) s vs 12 (10-13 [8-25]) s; p = 0.90), airway leak pressures (19 (16-21 [12-30]) cmH(2) O vs 18 (16-24 [10-34]) cmH(2) O; p = 0.55), fibreoptic position of the airway or drain tube, ease of gastric access and complications. Both devices provided effective ventilation requiring minimal airway manipulation. The LMA Supreme can be a useful alternative to the LMA ProSeal when single-use supraglottic devices with gastric access capabilities are required.
Medical subject headings
- Laryngeal Masks