Air-displacement plethysmography pediatric option in 2-6 years old using the four-compartment model as a criterion method.
Where this comes from
- Record sourced from PubMed, PMID 22421895.
- Also identified by DOI 10.1038/oby.2012.28 and PMC identifier 3628559.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The objective of this study was to determine the accuracy, precision, bias, and reliability of percent fat (%fat) determined by air-displacement plethysmography (ADP) with the pediatric option against the four-compartment model in 31 children (4.1 ± 1.2 years, 103.3 ± 10.2 cm, 17.5 ± 3.4 kg). %Fat was determined by (BOD POD Body Composition System; COSMED USA, Concord, CA) with the pediatric option. Total body water (TBW) was determined by isotope dilution ((2)H(2)O; 0.2 g/kg) while bone mineral was determined by dual-energy X-ray absorptiometry (DXA) (Lunar iDXA v13.31; GE, Fairfield, CT and analyzed using enCore 2010 software). The four-compartment model by Lohman was used as the criterion measure of %fat. The regression for %fat by ADP vs. %fat by the four-compartment model did not deviate from the line of identity where: y = 0.849(x) + 4.291. ADP explained 75.2% of the variance in %fat by the four-compartment model while the standard error of the estimate (SEE) was 2.09 %fat. The Bland-Altman analysis showed %fat by ADP did not exhibit any bias across the range of fatness (r = 0.04; P = 0.81). The reliability of ADP was assessed by the coefficient of variation (CV), within-subject SD, and Cronbach's α. The CV was 3.5%, within-subject SD was 0.9%, and Cronbach's α was 0.95. In conclusion, ADP with the pediatric option is accurate, precise, reliable, and without bias in estimating %fat in children 2-6 years old.
Medical subject headings
- Adipose Tissue
- Air
- Anthropometry
- Body Composition
- Plethysmography