The effect of retention of the femoral neck and of cement upon the stability of a proximal femoral prosthesis.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 2243216.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Radiological measurements of downward femoral component migration are reported for 203 hips in the first 2 years following total hip arthroplasty. The femora differed with respect to the retention of the femoral neck--in 167 it was retained and in 36 it was damaged--and to the use of cement--142 hips were press-fits and 61 were cemented. Multivariate analysis demonstrated that both retention of the neck and the use of cement retarded migration (ie, increased stability) of the femoral component (P = .0003 and .0037, respectively).
Medical subject headings
- Femur
- Hip Prosthesis
Anatomy
- femur
- hip