Chemical shift imaging of human brain: axial, sagittal, and coronal P-31 metabolite images.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2243963.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Multivoxel magnetic resonance (MR) spectroscopy and novel data analysis techniques were developed to obtain high-quality phosphorus-31 metabolite images from the human brain and to overlay each metabolite distribution directly onto corresponding hydrogen-1 MR images. The P-31 MR spectroscopic data were acquired by means of three-dimensional chemical shift imaging (phase encoding in three spatial dimensions) on a 1.5-T clinical instrument equipped with a specially designed quadrature P-31 birdcage coil constructed in the authors' laboratory. Axial, sagittal, and coronal metabolite images based on the area for any one of five peak regions (phosphodiester; phosphocreatine; gamma, alpha, and beta adenosine triphosphate) were generated from 8 X 8 X 8 or 12 X 12 X 8 CSI arrays with voxel sizes of 27 cm3 and 12 cm3, respectively. The positions of these images were aligned with anatomic features by means of the voxel-shifting capability of the Fourier transform. Direct overlays of these metabolite images on corresponding proton images demonstrated excellent correlation with anatomy, factors indicating the utility of this technique for viewing P-31 metabolite levels in all areas of the brain simultaneously.
Medical subject headings
- Brain
- Image Processing, Computer-Assisted
- Magnetic Resonance Imaging
- Magnetic Resonance Spectroscopy