Determination of left ventricular volume and mass with use of biphasic spin-echo MR imaging: comparison with cine MR.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 2243988.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this study, the authors compared a new rapid spin-echo magnetic resonance (MR) imaging method, biphasic MR, with cine MR in the determination of left ventricular volume and mass in healthy volunteers. Biphasic spin-echo MR images covering the entire heart were obtained with use of the electrocardiogram R wave and the downslope of the T wave at both end diastole and end systole, respectively. Biphasic MR-determined values correlated well with small standard errors of the estimate (end-diastolic volume = 7.82 cm3, end-diastolic mass = 10.20 g, end-systolic mass = 10.08 g, ejection fraction = 2.62%) and were more reproducible. Cine MR-defined end-systolic volume was significantly larger (P less than .01) and ejection fraction was significantly smaller (P less than .005) than biphasic MR-determined values probably because of the uncertainty in isolating end systole with cine MR. Left ventricular volumes, mass, and ejection fraction are more accurately and reproducibly quantified in a more time-efficient manner with use of biphasic MR than with cine MR because of its significantly shorter image acquisition and reconstruction times.
Medical subject headings
- Magnetic Resonance Imaging
- Stroke Volume
- Ventricular Function, Left