Use of plethysmographic variability index derived from the Massimo(®) pulse oximeter to predict fluid or preload responsiveness: a systematic review and meta-analysis.

Yin, J Y; Ho, K M · Anaesthesia · 2012

meta_analysis · Level I

Where this comes from

Abstract

This systematic review and meta-analysis assessed the accuracy of plethysmographic variability index derived from the Massimo(®) pulse oximeter to predict preload responsiveness in peri-operative and critically ill patients. A total of 10 studies were retrieved from the literature, involving 328 patients who met the selection criteria. Overall, the diagnostic odds ratio (16.0; 95% CI 5-48) and area under the summary receiver operating characteristic curve (0.87; 95% CI 0.78-0.95) for plethysmographic variability index to predict fluid or preload responsiveness was very good, but significant heterogeneity existed. This could be explained by a lower accuracy of plethysmographic variability index in spontaneously breathing or paediatric patients and those studies that used pre-load challenges other than colloid fluid. The results indicate specific directions for future studies.

Medical subject headings