The effect of riboflavin-ultraviolet A-induced collagen cross-linking on intraocular pressure measurement: an experimental study.

Livny, Eitan; Kaiserman, Igor; Hammel, Naama; Livnat, Tami; Zadok, David; Israel, Kremer; Bahar, Irit · Br J Ophthalmol · 2012

basic_science · Level V

Where this comes from

Abstract

To investigate the effect of corneal collagen cross-linking (CCL) on tonopen measurements of intraocular pressure (IOP). CCL with 0.1% riboflavin solution and 30 min of ultraviolet A radiation was performed on the right eye of 15 New Zealand albino adult rabbits (1.8-2.4 kg) (15 eyes). The left eye served as control. IOP was measured by a pressure transducer system (true IOP) and by the tonopen hand-held device (corneal applanation tonometer) before treatment, at 1 week, 1 month and 3 months following CCL. Reference pressure in the globe was increased by increments of 10 mm Hg from 10 to 40 mm Hg, using an anterior chamber infusion on a stand with variable height, and tonopen IOP measurements were recorded for each reference pressure in both eyes. Before CCL, tonopen readings were similar between the two eyes (p>0.05). Tonopen underestimated the true IOP in all cases. Following CCL treatment, IOP measurements were significantly higher in the treated eye, at all time intervals (0.005<p<0.03). The most significant difference between true and measured IOP was noted at a reference pressure of 20 mm Hg. IOP measurements following CCL are overestimated by the tonopen, probably due to increased stiffness of the treated cornea.