Modularity of the femoral component in total hip arthroplasty.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 22474091.
- Also identified by DOI 10.5435/JAAOS-20-04-214.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Modular femoral components have been developed to aid in recreating native femoral version, limb length, and offset in total hip arthroplasty. Use of modular implants results in cost savings, as well. Inventory can be reduced while allowing intraoperative flexibility and options. With modular implants, the femoral prosthesis can be built in situ, which is helpful in minimizing incision length and surgical dissection. However, additional modular junctions are associated with increased concern for component failure through taper fretting, fatigue fracture, and local corrosion, which may contribute to elevated serum metal ion levels. The recent trend toward using larger diameter femoral heads may impart higher loads and stress than were seen previously. Although modular components offer a plethora of intraoperative options in primary and revision total hip arthroplasty, the long-term effects of these additional junctions remains unknown.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Prosthesis Design
Anatomy
- femur
- hip