Failures of pin removal after in situ pinning of slipped capital femoral epiphyses: a comparison of different pin types.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 2250062.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We reviewed 75 pin retrievals after in situ pinning of 35 children with slipped capital femoral epiphysis (SCFE). Four types of pins were used. Ten pins in 8 hips could not be removed because of pin breakage or stripping. Our failure rate was significantly related to pin type and size (p less than 0.039). As compared with the expected values using the chi-square method of analysis, the noncannulated large steel pins and cannulated steel pin groups had fewer failures (p less than 0.001) as compared with the cannulated titanium and noncannulated small steel pin groups. We now avoid using cannulated titanium or noncannulated small pins in treatment of SCFE.
Medical subject headings
- Bone Nails
- Epiphyses, Slipped
- Femur Head
Anatomy
- femur