Daily orthogonal kilovoltage imaging using a gantry-mounted on-board imaging system results in a reduction in radiation therapy delivery errors.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 22503524.
- Also identified by DOI 10.1016/j.ijrobp.2012.01.033.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine whether the use of routine image guided radiation therapy (IGRT) using pretreatment on-board imaging (OBI) with orthogonal kilovoltage X-rays reduces treatment delivery errors. A retrospective review of documented treatment delivery errors from 2003 to 2009 was performed. Following implementation of IGRT in 2007, patients received daily OBI with orthogonal kV X-rays prior to treatment. The frequency of errors in the pre- and post-IGRT time frames was compared. Treatment errors (TEs) were classified as IGRT-preventable or non-IGRT-preventable. A total of 71,260 treatment fractions were delivered to 2764 patients. A total of 135 (0.19%) TEs occurred in 39 (1.4%) patients (3.2% in 2003, 1.1% in 2004, 2.5% in 2005, 2% in 2006, 0.86% in 2007, 0.24% in 2008, and 0.22% in 2009). In 2007, the TE rate decreased by >50% and has remained low (P = .00007, compared to before 2007). Errors were classified as being potentially preventable with IGRT (e.g., incorrect site, patient, or isocenter) vs. not. No patients had any IGRT-preventable TEs from 2007 to 2009, whereas there were 9 from 2003 to 2006 (1 in 2003, 2 in 2004, 2 in 2005, and 4 in 2006; P = .0058) before the implementation of IGRT. IGRT implementation has a patient safety benefit with a significant reduction in treatment delivery errors. As such, we recommend the use of IGRT in routine practice to complement existing quality assurance measures.
Medical subject headings
- Medical Errors
- Quality Improvement
- Radiotherapy, Image-Guided