Periprosthetic femoral fracture due to biodegradable cement restrictor.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 22521399.
- Also identified by DOI 10.1016/j.arth.2012.02.013.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Biodegradable materials are gaining popularity in orthopedics. Despite finding use in different areas of orthopedic surgery, they do not come without disadvantages such as foreign body reactions, granulomatous reactions, and sterile sinus formation in bone. We report a case of a patient who sustained a periprosthetic fracture seen at the tip of a cemented femoral stem approximately 5 years postsurgery, secondary to the use of a biodegradable cement restrictor. There was no evidence of trauma or fall on the affected hip. To our knowledge, there has been no previous report describing periprosthetic fracture at the tip of cemented femoral stem secondary to the use of a biodegradable cement restrictor. We suggest abandoning use of these materials while performing cemented hip arthroplasties.
Medical subject headings
- Absorbable Implants
- Arthroplasty, Replacement, Hip
- Bone Cements
- Femoral Fractures
- Hip Prosthesis
- Periprosthetic Fractures
Anatomy
- femur
- hip