Posterior femoral neck impingement secondary to excess acetabular anteversion in hip resurfacing arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 22560654.
- Also identified by DOI 10.1016/j.arth.2012.02.010.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Component malposition in resurfacing arthroplasty can cause impingement, reduced range of motion, increased metal wear, and early failure. It is therefore important that any component malposition is recognized early and monitored carefully. We present a case of delayed diagnosis of excess acetabular component anteversion causing posterior femoral neck impingement and pain. We correlate the degree of damage found intraoperatively with the degree of component malposition on the radiographs. This is used as a platform to discuss the optimum positioning of hip resurfacing components and the radiographic methods used to determine component position postoperatively.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femoracetabular Impingement
- Femur Neck
Anatomy
- femur
- hip