An injectable nucleus pulposus implant restores compressive range of motion in the ovine disc.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 22588378.
- Also identified by DOI 10.1097/BRS.0b013e31825cdfb7 and PMC identifier 3717581.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Investigation of injectable nucleus pulposus (NP) implant. To assess the ability of a recently developed injectable hydrogel implant to restore nondegenerative disc mechanics through support of NP functional mechanics. Although surgical intervention for low back pain is effective for some patients, treated discs undergo altered biomechanics and adjacent levels are at increased risk for accelerated degeneration. One potential treatment as an alternative to surgery for degenerated disc includes the percutaneous delivery of agents to support NP functional mechanics. The implants are delivered in a minimally invasive fashion, potentially on an outpatient basis, and do not preclude later surgical options. One of the challenges in designing such implants includes the need to match key NP mechanical behavior and mimic the role of native nondegenerate NP in spinal motion. The oxidized hyaluronic acid gelatin implant material was prepared. In vitro mechanical testing was performed in mature ovine bone-disc-bone units in 3 stages: intact, discectomy, and implantation versus sham. Tested samples were cut axially for qualitative structural observations. Discectomy increased axial range of motion (ROM) significantly compared with intact. Hydrogel implantation reduced ROM 17% (P < 0.05) compared with discectomy and returned ROM to intact levels (ROM intact 0.71 mm, discectomy 0.87 mm, postimplantation 0.72 mm). Although ROM for the hydrogel implant group was statistically unchanged compared with the intact disc, ROM for sham discs, which received a discectomy and no implant, was significantly increased compared with intact. The compression and tension stiffness were decreased with discectomy and remained unchanged for both implant and sham groups as expected because the annulus fibrosus was not repaired. Gross morphology images confirmed no ejection of NP implant. An injectable implant that mimics nondegenerate NP has the potential to return motion segment ROM to normal subsequent to injury.
Medical subject headings
- Intervertebral Disc
- Prostheses and Implants
- Range of Motion, Articular
Anatomy
- lumbar spine