Compressive strength of autologous and allogenous bone grafts for thoracolumbar and cervical spine fusion.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 2263975.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The selection of the bone graft type for stabilization of spinal fusion depends on availability, the clinical situation, and the desired mechanical stability. The authors determined the potential immediate postoperative compressive strength of various types of bone grafts under axial compression on a material testing machine. The fibular strut graft (5,070 +/- 3,250 N, mean +/- standard deviation [SD]) was significantly stronger (P less than 0.05) than the anterior (1,150 +/- 487 N) and posterior (667 +/- 311 N) iliac crest grafts, and the rib grafts (452 +/- 192 N). Hydroxyapatite grafts with a pore size of 200 mu were significantly stronger (P less than 0.05) than those with a pore size of 500 mu (1,420 +/- 480 N versus 338 +/- 78 N). Ethylenoxide sterilization had no significant effect on the immediate compressive strength. Bicortical and tricortical Bailey-Badgley and Cloward bone grafts also were compared. Results showed that all cervical graft types may be sufficiently strong to support sizable loads.
Medical subject headings
- Bone Transplantation
- Bone and Bones
- Cervical Vertebrae
- Spinal Fusion
Anatomy
- cervical spine
- lumbar spine