Repeated early failure of a newly designed hinged knee system.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 22749005.
- Also identified by DOI 10.1016/j.arth.2012.04.029.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a case of acute and recurrent accelerated wear of the polyethylene bushings of the hinge mechanism in the Zimmer Segmental Knee System. This resulted in an unacceptable recurvatum deformity leading to multiple revision knee operations, which was unexpected in a modern design hinged knee system. A custom modification of the original design was used to prevent further recurvatum deformity. The current design of the hinge post mechanism in this system appears to be inadequate, which led to the development of a significant recurvatum deformity in our patient. Surgeons should be aware of this potential complication when considering the use of this revision knee system.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Femoral Fractures
- Knee Joint
- Knee Prosthesis
- Periprosthetic Fractures
- Prosthesis Failure