Objective autorefraction in posterior chamber pseudophakia.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 2275935.
- Also identified by PMC identifier 1042277.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Automated refraction with the Canon RK-1 Autoref keratometer was evaluated in 110 eyes (110 patients) six to eight weeks after they had undergone extracapsular cataract extraction with posterior chamber intraocular lens implantation and achieved a best corrected visual acuity of at least 6/12. Autorefraction readings were obtained in 100 (91%) of these eyes. The agreement between autorefraction and clinical refraction data was 98% for spherical equivalence less than 0.51 dioptres (D), 95% for sphere power less than 0.51D, 94% for cylinder power less than 0.51D, and 85% for cylinder axis less than 11 degrees. Autorefraction can provide acceptably accurate postoperative refraction values in pseudophakic eyes.
Medical subject headings
- Lenses, Intraocular
- Refraction, Ocular