Computer-controlled dynamic phantom for ultrasound hyperthermia studies.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2276758.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A dynamic phantom system for use in evaluating hyperthermia heating equipment has been designed, constructed, and tested. It consists of four in vitro preserved canine kidneys perfused with an 80% ethanol preparation by a single pump, plus four computer-controlled valves and four flow meters. The flow rate to each kidney is computer-controlled, giving flexibility in the types of spatial and temporal flow patterns possible. Examples are given for both step and ramp changes in flow, and for a temperature dependent flow which has been used to simulate physiological responses to elevated temperatures.
Medical subject headings
- Hyperthermia, Induced
- Kidney
- Models, Biological
- Models, Structural