Electrical stimulation with Pt electrodes. VIII. Electrochemically safe charge injection limits with 0.2 ms pulses.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2276759.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The charge injection limits of a Pt electrode using 0.2 ms charge balanced, biphasic current pulses ranged from 50 to 150 microC/cm2 geometric if the potential excursions of the electrode are kept below those at which H2 or O2 is produced. These charge densities are three to ten times smaller than the currently accepted value based on earlier experiments in which the reversible surface reactions were fully utilized and the pulse widths were longer.
Medical subject headings
- Biocompatible Materials
- Electric Stimulation
- Electrodes
- Platinum