What are the predictors for bead shedding in porous-coated hip and knee arthroplasties?: An implant retrieval study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 22770858.
- Also identified by DOI 10.1016/j.arth.2012.05.020.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Bead shedding has been reported in both hip and knee arthroplasties and can result in increased polyethylene wear and metal ion release. The rates of bead loss were recorded for each device type, and the interfacial bead-substrate shear load to failure and contact areas of the beads were determined. The 3 components with the highest rate of bead loss, namely, S-Rom, Omnifit, and Optifix acetabular shells, recorded the lowest shear loads to failure. Percentage bead-substrate contact area ranged between 7% and 37% and was the most important factor affecting bead shedding. Other factors influencing bead loss in decreasing order of importance were bead fusion, bead size, component type (acetabular or knee), component loosening, and time in situ.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Arthroplasty, Replacement, Knee
Anatomy
- hip
- knee