Nasal pulse oximetry overestimates oxygen saturation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 2278333.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Ten surgical patients were monitored with nasal and finger pulse oximetry (Nellcor N-200) for five study periods with alternating mouth and nasal breathing and switching of cables and sensors. Nasal pulse oximetry was found to overestimate arterial oxygen saturation by 4.7 (SD 1.4%) (bias and precision).