Epidural catheter connectors: a laboratory-based comparison of the Portex Tuohy-Borst and EpiFuse™ designs.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 22804723.
- Also identified by DOI 10.1111/j.1365-2044.2012.07248.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Disconnection of an epidural catheter from its connector may result in patient harm and commonly requires resiting of the epidural. Clamp-connector designs such as the novel Portex EpiFuse™ potentially offer an improved safety profile over screw-cap designs such as the Tuohy-Borst, but comparative studies are limited. We therefore compared the tensile strength of EpiFuse and Tuohy-Borst connectors in a laboratory setting. We further sought to establish whether operator modification of the EpiFuse increased its vulnerability to disconnection. The median (IQR [range]) force to induce disconnection was 8.0 (4.1-12.8 [0.0-22.6]) N for Tuohy-Borst connectors and 16.4 (15.2-17.7 [5.7-18.9]) and 15.9 (15.0-16.9 [5.8-18.1]) N for standard and modified EpiFuse connectors, respectively (p<0.0001). The Tuohy-Borst was also less likely to meet British Standard requirements (13/20 sets vs 19/20 and 20/20, p=0.002). Modification of the EpiFuse did not affect lumen patency or connection strength. We conclude that under controlled conditions, EpiFuse connectors are superior to Tuohy-Borst connectors.
Medical subject headings
- Anesthesia, Epidural
- Catheters