Severe bending of two aortic stent-grafts: an experimental and numerical mechanical analysis.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 22805981.
- Also identified by DOI 10.1007/s10439-012-0618-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Stent-grafts (SGs) are commonly used for treating abdominal aortic aneurysms (AAAs) and numerical models tend to be developed for predicting the biomechanical behavior of these devices. However, due to the complexity of SGs, it is important to validate the models. In this work, a validation of the numerical model developed in Demanget et al. (J. Mech. Behav. Biomed. Mater. 5:272-282, 2012) is presented. Two commercially available SGs were subjected to severe bending tests and their 3D geometries in undeformed and bent configurations were imaged from X-ray microtomography. Dedicated image processing subroutines were used in order to extract the stent centerlines from the 3D images. These skeletons in the undeformed configurations were used to set up SG numerical models that are subjected to the boundary conditions measured experimentally. Skeletons of imaged and deformed stents were then quantitatively compared to the numerical simulations. A good agreement is found between experiments and simulations. This validation offers promising perspectives to implementing the numerical models in a computer-aided tool and simulating the endovascular treatments.
Medical subject headings
- Aorta
- Computer Simulation
- Equipment Failure Analysis
- Models, Cardiovascular
- Stents