Accuracy and speed of robotic assisted needle interventions using a modern cone beam computed tomography intervention suite: a phantom study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 22821395.
- Also identified by DOI 10.1007/s00330-012-2585-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To analyse the feasibility and accuracy of robotic aided interventions on a phantom when using a modern C-arm-mounted cone beam computed tomography (CBCT) device in combination with needle guidance software. A small robotic device capable of holding and guiding needles was attached to the intervention table. After acquiring a 3D data set the access path was planned on the CBCT workstation and shown on the intervention monitor. Then the robot was aligned to the live fluorosopic image. A total of 40 punctures were randomly conducted on a phantom armed with several targets (diameter 2 mm) in single and double oblique trajectory (n = 20 each). Target distance, needle deviation and time for the procedures were analysed. All phantom interventions (n = 40) could be performed successfully. Mean target access path within the phantom was 8.5 cm (min 4.2 cm, max 13.5 cm). Average needle tip deviation was 1.1 mm (min 0 mm, max 4.5 mm), time duration was 3:59 min (min 2:07 min, max 10:37 min). When using the proposed robot device in a CBCT intervention suite, highly accurate needle-based interventional punctures are possible in a reasonable timely manner in single as well as in double oblique trajectories.
Medical subject headings
- Biopsy, Needle
- Cone-Beam Computed Tomography
- Phantoms, Imaging
- Robotics