SLAP tears: diagnosis using 3-T shoulder MR arthrography with the 3D isotropic turbo spin-echo space sequence versus conventional 2D sequences.

Jung, Joon-Yong; Jee, Won-Hee; Park, Michael Yong; Lee, So-Yeon; Kim, Yang-Soo · Eur Radiol · 2013

retrospective_cohort · Level III

Where this comes from

Abstract

The aim of this study was to determine the accuracy and reliability of shoulder magnetic resonance (MR) arthrography with three-dimensional (3D) isotropic intermediate-weighted turbo spin-echo (TSE) sampling perfection with application-optimised contrasts using different flip angle evolution (SPACE) in the diagnosis of superior labrum anterior-to-posterior (SLAP) lesions compared with two-dimensional (2D) TSE at 3.0 T. MR arthrograms, including 2D TSE and 3D TSE-SPACE, in 87 patients who underwent arthroscopy were retrospectively analysed by two reviewers for the presence and type of SLAP lesions. Sensitivity and specificity were compared using McNemar's test, and inter-observer agreement was calculated using Cohen's kappa. Receiver operating characteristic (ROC) curve analyses were performed. The mean sensitivity, specificity and accuracy were 90%, 85% and 86% for 2D TSE, and 81%, 86% and 85% for 3D TSE-SPACE respectively, with no statistically significant differences. Inter-observer agreements were substantial in 2D TSE (κ = 0.76) and 3D TSE-SPACE (κ = 0.68). The areas under the ROC curves were 0.92 for 2D TSE and 0.90 for 3D TSE-SPACE, which were not significantly different. MR arthrography with 3D TSE-SPACE showed comparable accuracy and substantial inter-observer agreement for the diagnosis of SLAP lesions.

Medical subject headings