Incidence of osteolysis in total knee arthroplasty: comparison between radiographic and retrieval analysis.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 22868074.
- Also identified by DOI 10.1016/j.arth.2012.06.008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Detection of polyethylene wear and osteolysis in total knee arthroplasty using radiographs is imprecise. To correlate radiographic findings with retrieved tibial component analysis, we reviewed knee revision cases. A joint registry was used to identify all knee revisions. Radiographic scoring systems were developed. Radiographs were analyzed for osteolysis and implant alignment. Polyethylene liner retrievals were visually and optically graded for surface damage. Statistical analyses that included correlation analysis and Mann-Whitney U test were used. When osteolysis was found, radiographic scores and delamination score were significantly higher (P = .047 and P = .014, respectively). Delamination is a good indicator for polyethylene wear and osteolysis. There is a need to develop in vivo methods to identify polyethylene changes and thereby prevent severe clinical outcomes.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Joint
- Knee Prosthesis
- Osteolysis
Anatomy
- knee