Abstract
The two-piece Van Straten Leuwen Poeschmann Metal (LPM) prosthesis was intended for the proximal interphalangeal joints. However, revision rates of 29% after 19 months were reported, as well as massive osteolysis. Five failed LPM titanium-niobium coated cobalt chromium components were obtained, three distal and two proximal, and subjected to a forensic retrieval analysis. Components were analyzed using a Talysurf contacting profilometer, ZYGO noncontacting profilometer, and environmental-scanning electron microscope. All components were heavily worn. In some regions the titanium-niobium coating had been scratched and penetrated. Elsewhere this coating had been removed where there was minimal scratching, which may have been due to corrosion between the coating and substrate. The osteolysis reported clinically was likely to be linked to the wear debris from the failed titanium-niobium coating and substrate.
Medical subject headings
- Equipment Failure Analysis
- Finger Joint
- Prosthesis Failure
Anatomy
- hand